HelpContactFor teachers and institutes
>_StemPy Lab
Mock TestsClasses

>_StemPy Lab

Learn. Solve. Code the solutions. Test yourself.

Live examsFree mock exams, timed and rankedEveryone starts together, and your score and rank come the moment it ends.See the exams →
Live examsInternational live exams are comingTell us which exam you want first. Timed practice papers are open now.See what is coming →

Learn

  • Physics
  • Chemistry
  • Mathematics
  • Biology
  • Python Programming
  • Statistics & Data Science
  • Computer Science
  • Space Science and Astronomy
  • Classes
  • Groups

Practice

  • Problems
  • Quizzes
  • Mock Tests
  • Prepare for an Exam
  • Question Sets
  • Exams
  • Certificates
  • Python Labs
  • AI Practice Generator
  • Student-Created Tests

Community

  • Discussions
  • Resources
  • Notes
  • Glossary
  • Blog

Account

  • Log in
© 2026 senLearn. All rights reserved.HelpFor SchoolsAboutContactPrivacy PolicyTerms of UseRefund Policy
  1. Problems
  2. /A tube gives a short exposure of 30 kW for 0.10 s…
AdvancedPhysicsHigh School (11–12)Medical PhysicsNumericalConceptualA-Level

A tube gives a short exposure of 30 kW for 0.10 s…

StemPy LabContributed by StemPy Lab

Try it on paper first. A hint, the key idea and the worked solution are here when you need them.

A tube gives a short exposure of 303030 kW for 0.100.100.10 s. About 99 %99\ \%99 % of the energy becomes heat in a tungsten anode disc of mass 0.500.500.50 kg, with a specific heat capacity of 134 J kg−1K−1134\ \text{J kg}^{-1}\text{K}^{-1}134 J kg−1K−1. Find the heat and the rise in temperature if the heat spreads through the whole disc. To make a characteristic line, an electron must first knock out an electron of the innermost shell of tungsten, which needs 69.569.569.5 keV. The strongest line has a photon energy of 59.359.359.3 keV. Find the least tube voltage that gives the line, and the wavelength of the line. Why does the anode spin?

Check your answer

Work it out on paper, then type your number. The answer itself stays hidden.

J
K
kV
pm
Need a reminder? The key ideaShowHide

An electron that crosses a voltage VVV arrives with the energy eVeVeV. The most energetic photon takes all of it, so

λmin=hceV\displaystyle \lambda_{min} = \frac{hc}{eV}λmin​=eVhc​

with hc/e=1239.8hc/e = 1239.8hc/e=1239.8 nm V. The power of the beam of electrons is P=IVP = IVP=IV, and only about 1 %1\ \%1 % of it becomes X-rays.

The worked solution

Check an answer or open a hint to unlock the worked solution.

Have you tried it?

Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.

Step 1. The heat. E=30 000 W×0.10 s=3000E = 30\,000\ \text{W} \times 0.10\ \text{s} = 3000E=30000 W×0.10 s=3000 J, and 99 %99\ \%99 % of it is 297029702970 J.

Step 2. The rise. We want ΔT\Delta TΔT, so ΔT=Qmc=2970 J0.50 kg×134 J kg−1K−1=44.3\Delta T = \dfrac{Q}{mc} = \dfrac{2970\ \text{J}}{0.50\ \text{kg} \times 134\ \text{J kg}^{-1}\text{K}^{-1}} = 44.3ΔT=mcQ​=0.50 kg×134 J kg−1K−12970 J​=44.3 K.

Step 3. The least voltage. Each electron brings eVeVeV, so V=69.5V = 69.5V=69.5 kV at least.

Step 4. The line. λ=1239.8 eV nm59 300 eV=0.0209\lambda = \dfrac{1239.8\ \text{eV nm}}{59\,300\ \text{eV}} = 0.0209λ=59300 eV1239.8 eV nm​=0.0209 nm, which is 20.920.920.9 pm.

Answer: 297029702970 J, 44.344.344.3 K, 69.569.569.5 kV and 20.920.920.9 pm.

Why it spins. The heat does not spread through the whole disc at once. It lands on a focal spot of a few square millimetres. A spinning anode brings a fresh part of a ring under the beam all the time, so no one spot melts. The line has less energy than 69.569.569.5 keV, because the electron that fills the gap comes from a shell that is itself bound.

Log in to save this for later
← All Problems
← A radiographer changes a tube from 60 kV and 10 mA to 90 kV and 5.0 mA, to image a thicker part of…A beam holds soft photons of 20 keV and harder ones of 60 keV… →
More onMedical Physics

Discussion (0)

Log in to join the discussion.

Loading…

Learn it in

  1. Making X-rays: the tube and its spectrumX-Ray Imaging, CT and MRI

More on this topic

  1. A tumour of mass 0.12 kg is to receive 60 Gy in 30 equal fractions…Advanced
  2. After a large dose of iodine-131, a patient gives a dose rate of 25 microsievert per hour at 1.0 m…Hard
  3. At the start of a scan, a patient holds 300 MBq of fluorine-18, whose half-life is 109.8 min…Advanced
A cyclotron makes fluorine-18, with a half-life of 109.8 min, for a hospital 2.0 h away…Hard
  • A PET ring has a diameter of 80 cm. For one event, the photon that reaches detector A arrives 300…Hard
  • All problems on this topic →