After a large dose of iodine-131, a patient gives a dose rate of microsievert per hour at m. Treat the patient as a small source. Find the dose rate at m and at m. A visitor sits at m and may receive at most mSv. For how many hours may she stay, if the dose rate stayed the same? A lead screen with a half-value thickness of mm is mm thick. What is the dose rate at m behind it?
Absorbed dose in gray. Equivalent dose in sievert. Effective dose . From a small source the dose rate falls as , and each half-value thickness of shielding halves it.
Check an answer or open a hint to unlock the worked solution.
Have you tried it?
Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.
Step 1. At 1.5 m. .
Step 2. At 2.5 m. .
Step 3. The time. h.
Step 4. Behind the lead. .
Answer: , , h and .
Check. In fact the rate falls each day as the iodine decays and leaves the body, so the true limit is even longer. Distance alone cuts the rate by more than six times, before any lead.