Absorbed dose is energy per kilogram, in gray. The equivalent dose, in sievert, weights it for the kind of radiation, with alpha particles 20 times as harmful as gamma rays. Radiotherapy aims a large dose at a tumour from several directions, in small daily fractions, so that healthy tissue can recover.
about 30 min1 simulation
Lesson 3 of 3Show all lessonsHide
Try the code and the simulations beside the lesson.
Radiation that harms, and radiation that heals
The same ionising radiation that lets us see inside the body can damage it. It breaks bonds in DNA, and a cell may die or, more rarely, turn into a cancer cell years later. So every scan is a trade between what we learn and the small risk. And in radiotherapy we use that very damage, on purpose, to kill a tumour.
To weigh this, we need to measure the dose.
Absorbed dose: the gray
The absorbed doseD is the energy E absorbed by each kilogram of tissue:
D=mE
Its unit is the gray (Gy), with 1Gy=1J/kg. A tumour of 0.050 kg that absorbs 0.10 J has received D=0.050kg0.10J=2.0 Gy.
Equivalent dose: the sievert
The same absorbed dose of different radiations does not do the same harm. Alpha particles pack their ionisation into a very short track, so they damage DNA far more. We multiply by a radiation weighting factorwR:
H=wRD
The equivalent doseH is in sievert (Sv). The factors used today, from ICRP Publication 103, are 1 for photons and electrons, 2 for protons and 20 for alpha particles. Neutrons have factors from 2.5 to 20, by their energy.
A bar chart of the radiation weighting factors used to turn absorbed dose in gray into equivalent dose in sievert. Gamma rays, X-rays and beta particles each have a bar of height 1. Protons have a bar of height 2. Alpha particles have a tall bar of height 20. A second small table lists tissue weighting factors: 0.12 for the lungs, stomach, colon, red bone marrow and breast, 0.08 for the gonads, 0.04 for the thyroid, liver, bladder and oesophagus, and 0.01 for skin, bone surface, brain and salivary glands.
Effective dose: the whole body
Organs differ in how easily radiation causes cancer in them. The effective doseE adds the equivalent dose of each tissue, times its tissue weighting factorwT:
E=∑wTHT
The tissue factors add up to 1. Some of them are 0.12 for the lungs, the stomach and the red bone marrow, 0.04 for the thyroid and 0.01 for the skin. So the effective dose is the dose to the whole body that would carry the same risk.
Worked example. A scan gives 2.0 mSv to the lungs, 0.5 mSv to the stomach, 1.5 mSv to the thyroid and 4.0 mSv to the skin, and nothing else. Then E=0.12×2.0+0.12×0.5+0.04×1.5+0.01×4.0, all in mSv. So E=0.24+0.06+0.06+0.04=0.40 mSv.
How big are these doses?
Everyone receives a background dose from rocks, radon gas, cosmic rays and food. The world average is about 2.4 mSv in a year, and it varies a great deal from place to place. Some rough doses from medicine:
Procedure
Effective dose
X-ray of the chest
about 0.02 mSv
CT of the abdomen
about 10 mSv
PET with CT
about 10 to 25 mSv
Two kinds of effect. At high doses, above about 1 Gy at once, cells die in large numbers. Burns and radiation sickness follow, and they appear only above a threshold. These are deterministic effects. At low doses the only effect is a small chance of cancer later. This is a stochastic effect. For radiation protection we assume the risk is proportional to the dose, about 5% for each sievert, with no safe threshold.
Protecting staff and patients
The rule is to keep doses as low as reasonably achievable. Three tools do it.
Time. Half the time near a source gives half the dose.
Distance. From a small source the dose rate falls as 1/r2. Doubling the distance gives a quarter of the dose.
Shielding. Each half-value thickness of lead or concrete halves the dose rate.
Radiotherapy
A tumour needs a far larger dose than any scan, often 60 Gy or more in all. How do we give it without killing the healthy tissue around?
Many beams. A linear accelerator makes X-rays of several megavolt. The beam is aimed at the tumour from several directions. The tumour gets every beam. Healthy tissue on each path gets only one. This is called cross-fire.
Fractionation. The dose is split into small daily fractions, for example 2 Gy a day, five days a week, for six weeks. Healthy cells repair between fractions better than tumour cells do.
Shaping. Leaves of tungsten shape each beam to the outline of the tumour.
A cross-section of a body with a tumour near the centre. Three beams of X-rays from a linear accelerator enter from three directions and cross at the tumour. Each beam is drawn as a shaded band. Where all three bands overlap, at the tumour, the shading is darkest and is labelled three beams, full dose. Along each single band through healthy tissue the shading is light and labelled one beam, one third of the dose.
Other methods carry the source to the tumour. In brachytherapy, small sealed sources are placed inside or next to the tumour. In iodine therapy, the thyroid takes up iodine-131, and its beta particles destroy the gland from inside. Proton beams deposit most of their energy at the end of their range, in the Bragg peak. By choosing their energy, the peak is placed in the tumour, with little dose beyond it.
Start at 1 m and step back to 2 m. By what factor does the rate fall? Now add two half-value thicknesses of lead. What is the rate at 2 m now?
Practise with new numbers
Try it in Python
The program finds the equivalent dose in four tissues and the effective dose, and then the dose rate at four distances from a small source.
Common mistakes
Mixing up gray and sievert. The gray is energy per kilogram. The sievert weights it for the kind of radiation and, for the effective dose, the tissue.
Using 1/r for distance. From a small source the dose rate falls as 1/r2.
Thinking a dose of 2 Gy heats a tumour. It warms the tissue by less than a thousandth of a kelvin. The harm comes from ionisation, not heat.
Saying deterministic effects can follow any dose. They need a dose above a threshold. Below it, the risk is stochastic.
Remember these
D=E/m in gray. H=wRD in sievert, with wR=1 for photons, 2 for protons, 20 for alpha.
E=∑wTHT, and the wT add up to 1.
Background about 2.4 mSv a year. Risk about 5% per sievert.