In PET a tracer such as fluorine-18 gives out positrons. Each positron meets an electron within a millimetre or so, and the pair turns into two gamma photons of 0.511 MeV that fly apart in opposite directions. A ring of detectors that catches both photons at once finds a line through the event.
about 25 min1 simulation
Lesson 2 of 3Show all lessonsHide
Try the code and the simulations beside the lesson.
A tracer that sends out antimatter
Cancer cells grow fast, so they use a lot of glucose. Suppose we could give a patient sugar that glows wherever it is used. A tumour, or a busy part of the brain, would light up. Positron emission tomography (PET) does exactly this.
Fluorine-18 and FDG
The usual tracer is fluorodeoxyglucose (FDG), a form of glucose in which one atom is fluorine-18. Cells take it up as they take up glucose. Fluorine-18 decays by beta-plus decay, with a half-life of 110 min:
918F→818O
plus a positron e+ and a neutrino ν.
Its half-life is short, so it is made in a cyclotron close to the hospital, and used within hours.
Annihilation
The positron is the antiparticle of the electron. In tissue it travels about a millimetre and slows down. Then it meets an electron and the two annihilate: both vanish and their mass becomes energy.
Energy. Both are almost at rest, so the energy is the two rest energies, 2mec2.
Momentum. The total momentum was nearly zero, so it must stay zero. One photon cannot do that. So two photons fly off in opposite directions, each with the same energy mec2.
Eγ=mec2
With me=9.109×10−31 kg and c=2.998×108 m/s, Eγ=8.19×10−14 J. In MeV that is 1.602×10−13J/MeV8.19×10−14J=0.511 MeV. The wavelength of each photon is λ=hc/Eγ=2.43 pm.
The PET ring
A ring of detectors around the head of a patient, seen end on. Inside the brain a positron meets an electron, and two gamma photons of 0.511 megaelectronvolt each fly off in opposite directions along one straight line to two detectors on opposite sides of the ring. The line between the two detectors is drawn dashed and labelled line of response. A small clock symbol at each detector shows that the two arrivals are compared in time.
The patient lies inside a ring of detectors, each a scintillator crystal with a photomultiplier.
Coincidence. Two detectors that each catch a 0.511 MeV photon within a few nanoseconds of each other are taken to have seen one annihilation.
The line of response. The event lies somewhere on the straight line between those two detectors.
Many lines. Millions of such lines cross where the tracer is most concentrated. A computer turns them into a picture of each slice, as in CT.
Time of flight. Modern detectors can also measure which photon arrived first. Suppose the event is a distance x from the midpoint of the line. One photon travels x farther than the other, so the paths differ by 2x. Then cΔt=2x, and we want x:
x=2cΔt
Take a difference of 400 ps:
x=21×2.998×108m/s
×4.0×10−10s
x=0.060m
So the event is 6.0 cm from the midpoint. A real detector times to a few hundred picoseconds, so this only narrows the place along the line. It still makes the picture clearer.
PET with CT
A PET picture shows where the tracer went, but not the shape of the organs around it. So PET is nearly always combined with a CT scan in the same machine. The CT shows the anatomy, and the PET shows the activity on top of it. A doctor can then see that a bright spot is, for example, in a lymph node next to the lung.
Set the difference to 0. Where was the event? Now set it to −400 ps. Which way did the event move, and how far?
Practise with new numbers
Try it in Python
The program finds the energy and the wavelength of each photon, and the place of an event for four differences in arrival time.
Common mistakes
Saying one photon is made. One photon cannot carry away zero momentum. Two photons go in opposite directions.
Using 2mec2 for each photon. The total is 2mec2, so each photon has mec2=0.511 MeV.
Using x=cΔt. The paths differ by twice the distance from the midpoint.
Thinking the camera sees the positron. It sees the two gamma photons from the annihilation.
Remember these
FDG carries fluorine-18 (110 min) to cells that use glucose fast.
e++e−→2γ, each photon 0.511 MeV, in opposite directions.
Coincidence in a ring of detectors gives a line of response.
Time of flight: x=cΔt/2 from the midpoint.
PET shows activity. CT, in the same machine, shows the anatomy.
Check yourself
Try each one on your own first. If you are stuck, open a hint. Then check with Show the answer.
Why must the two photons of an annihilation at rest travel in opposite directions?
Two photons reach the ring at the same moment. Where was the annihilation?
Why is fluorine-18 made close to the hospital?
Why do PET scanners also have a CT scanner built in?
References
OpenStax College Physics 2e, Chapter 32 Medical Applications of Nuclear Physics, section 32.1, on PET.
OpenStax University Physics Volume 3, Chapter 11 Particle Physics and Cosmology, on antiparticles and annihilation.