Try it on paper first. A hint, the key idea and the worked solution are here when you need them.
The face of an ultrasound probe has an acoustic impedance of . A coupling gel has and the skin has . Without the gel, a thin layer of air with sits between the probe and the skin. Ignore attenuation, and count only the first crossing of each boundary. Find the share of the intensity that enters the skin with the gel and without it. The echo must cross the same boundaries on the way back. By what factor, and by how many decibel, is a returning echo weaker without the gel?
The acoustic impedance of a material is . At a boundary met at right angles, the share of the intensity reflected is
and the share passed on is .
Check an answer or open a hint to unlock the worked solution.
Have you tried it?
Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.
Plan. Multiply the shares passed on at the boundaries on the way in. Then square, for the way in and the way out.
Step 1. With the gel. Probe to gel: . Gel to skin: . The product is .
Step 2. Through air. Probe to air: . Air to skin: . The product is .
Step 3. The echo. The ratio for one way is . Both ways, it is . In decibel, dB.
Answer: with the gel, through air, and the echo is times, or dB, weaker without the gel.
Check. No amplifier can make up dB. A probe on dry skin shows a dark screen, as every sonographer knows.