HelpContactFor teachers and institutes
>_StemPy Lab
Mock TestsClasses

>_StemPy Lab

Learn. Solve. Code the solutions. Test yourself.

Live examsFree mock exams, timed and rankedEveryone starts together, and your score and rank come the moment it ends.See the exams →
Live examsInternational live exams are comingTell us which exam you want first. Timed practice papers are open now.See what is coming →

Learn

  • Physics
  • Chemistry
  • Mathematics
  • Biology
  • Python Programming
  • Statistics & Data Science
  • Computer Science
  • Space Science and Astronomy
  • Classes
  • Groups

Practice

  • Problems
  • Quizzes
  • Mock Tests
  • Prepare for an Exam
  • Question Sets
  • Exams
  • Certificates
  • Python Labs
  • AI Practice Generator
  • Student-Created Tests

Community

  • Discussions
  • Resources
  • Notes
  • Glossary
  • Blog

Account

  • Log in
© 2026 senLearn. All rights reserved.HelpFor SchoolsAboutContactPrivacy PolicyTerms of UseRefund Policy
  1. Problems
  2. /A piezoelectric disc in a probe is 0.50 mm thick…
HardPhysicsHigh School (11–12)Medical PhysicsNumericalA-Level

A piezoelectric disc in a probe is 0.50 mm thick…

StemPy LabContributed by StemPy Lab

Try it on paper first. A hint, the key idea and the worked solution are here when you need them.

A piezoelectric disc in a probe is 0.500.500.50 mm thick. Sound travels at 400040004000 m/s inside the crystal. The disc resonates when its thickness is half a wavelength of sound in the crystal. Find its resonant frequency. Find the wavelength of the ultrasound in soft tissue, where c=1540c = 1540c=1540 m/s. Each pulse lasts three cycles. Two boundaries along the beam give separate echoes only if they are farther apart than half the length of the pulse in the tissue. Find that smallest separation.

Check your answer

Work it out on paper, then type your number. The answer itself stays hidden.

MHz
mm
mm
Need a reminder? The key ideaShowHide

The acoustic impedance of a material is Z=ρcZ = \rho cZ=ρc. At a boundary met at right angles, the share of the intensity reflected is

α=(Z2−Z1)2(Z2+Z1)2\displaystyle \alpha = \frac{(Z_2 - Z_1)^2}{(Z_2 + Z_1)^2}α=(Z2​+Z1​)2(Z2​−Z1​)2​

and the share passed on is 1−α=4Z1Z2(Z1+Z2)21 - \alpha = \dfrac{4Z_1Z_2}{(Z_1 + Z_2)^2}1−α=(Z1​+Z2​)24Z1​Z2​​.

The worked solution

Check an answer or open a hint to unlock the worked solution.

Have you tried it?

Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.

Step 1. In the crystal. λc=2×0.50 mm=1.0\lambda_c = 2 \times 0.50\ \text{mm} = 1.0λc​=2×0.50 mm=1.0 mm. So f=4000 m/s1.0×10−3 m=4.0×106f = \dfrac{4000\ \text{m/s}}{1.0 \times 10^{-3}\ \text{m}} = 4.0 \times 10^6f=1.0×10−3 m4000 m/s​=4.0×106 Hz, which is 4.04.04.0 MHz.

Step 2. In the tissue. The frequency does not change. λ=1540 m/s4.0×106 Hz=3.85×10−4\lambda = \dfrac{1540\ \text{m/s}}{4.0 \times 10^6\ \text{Hz}} = 3.85 \times 10^{-4}λ=4.0×106 Hz1540 m/s​=3.85×10−4 m, which is 0.3850.3850.385 mm.

Step 3. The pulse. Three cycles are 3×0.385 mm=1.1553 \times 0.385\ \text{mm} = 1.1553×0.385 mm=1.155 mm long. The second echo travels an extra path of twice the separation sss. The echoes are separate when 2s2s2s is more than the pulse length, so s=1.155 mm2=0.578s = \dfrac{1.155\ \text{mm}}{2} = 0.578s=21.155 mm​=0.578 mm.

Answer: 4.04.04.0 MHz, 0.3850.3850.385 mm and 0.5780.5780.578 mm.

What if? A thinner disc gives a higher frequency, a shorter pulse and finer detail along the beam. It also gives less depth.

Log in to save this for later
← All Problems
← The face of an ultrasound probe has an acoustic impedance of 1.70 × 10⁶ kg m⁻²s⁻¹…A beam goes into the body through a layer of fat 2.0 cm thick, then a layer of muscle 3.0 cm thick… →
More onMedical Physics

Discussion (0)

Log in to join the discussion.

Loading…

Learn it in

  1. Ultrasound: transducers and acoustic impedanceUltrasound, Hearing and Sight

More on this topic

  1. A tumour of mass 0.12 kg is to receive 60 Gy in 30 equal fractions…Advanced
  2. After a large dose of iodine-131, a patient gives a dose rate of 25 microsievert per hour at 1.0 m…Hard
  3. At the start of a scan, a patient holds 300 MBq of fluorine-18, whose half-life is 109.8 min…Advanced
A cyclotron makes fluorine-18, with a half-life of 109.8 min, for a hospital 2.0 h away…Hard