Try it on paper first. A hint, the key idea and the worked solution are here when you need them.
Iron(II) ions and 1,10-phenanthroline give an orange red complex that absorbs strongly at 510 nm. A chemist makes five standards with 1.00, 2.00, 3.00, 4.00 and 5.00 mg/dm³ of iron, and reads absorbances of 0.198, 0.401, 0.597, 0.802 and 0.999. She dissolves one iron tablet, reduces all the iron to iron(II) and makes the solution up to 250.0 cm³. She takes 2.00 cm³ of it, adds the reagents and makes it up to 100.0 cm³. This reads . Find the slope and the intercept of the least-squares calibration line, the iron in the measured solution, and the mass of iron in the tablet.
Check an answer or open a hint to unlock the worked solution.
Have you tried it?
Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.
Step 1. The fit. For the five points, the least-squares line is , with in mg/dm³.
Step 2. The measured solution. We want , so .
Step 3. Undo the dilution. went to , a factor of 50. The flask held .
Step 4. The tablet. of iron.
Answer: slope per mg/dm³, intercept , and .
Check. The reading 0.655 lies between the third and fourth standards, inside the calibrated range. The intercept is close to zero, so the blank was set well.
Log in to join the discussion.
Loading…
Absorbance , and the Beer-Lambert law is . A calibration line of against gives the concentration of an unknown. With a fraction of stray light, the meter reads .