HelpContactFor teachers and institutes
>_StemPy Lab
Mock TestsClasses

>_StemPy Lab

Learn. Solve. Code the solutions. Test yourself.

Live examsFree mock exams, timed and rankedEveryone starts together, and your score and rank come the moment it ends.See the exams →
Live examsInternational live exams are comingTell us which exam you want first. Timed practice papers are open now.See what is coming →

Learn

  • Physics
  • Chemistry
  • Mathematics
  • Biology
  • Python Programming
  • Statistics & Data Science
  • Computer Science
  • Space Science and Astronomy
  • Classes
  • Groups

Practice

  • Problems
  • Quizzes
  • Mock Tests
  • Prepare for an Exam
  • Question Sets
  • Exams
  • Certificates
  • Python Labs
  • AI Practice Generator
  • Student-Created Tests

Community

  • Discussions
  • Resources
  • Notes
  • Glossary
  • Blog

Account

  • Log in
© 2026 senLearn. All rights reserved.HelpFor SchoolsAboutContactPrivacy PolicyTerms of UseRefund Policy
Colorimetry and Titration Technique: Concentration and Its Uncertainty
  1. 1Colorimetry: the Beer-Lambert law at the bench
  2. 2Colorimetry at work: rates, formulas of complexes and equilibrium
  3. 3Titration technique, uncertainty and error
  1. Chemistry
  2. /Colorimetry and Titration Technique: Concentration and Its Uncertainty
  3. /Colorimetry: the Beer-Lambert law at the bench

Course 40.2·Lesson 1 of 3

Colorimetry: the Beer-Lambert law at the bench

Hold a glass of weak tea and a glass of strong tea against a window. Less light comes through the strong one. A colorimeter turns that difference into a number. This lesson explains absorbance, the Beer-Lambert law, how a colorimeter is set up, and how a calibration line gives the concentration of an unknown.

about 25 min1 simulation1 practice question
Lesson 1 of 3Show all lessonsHide
Colorimetry and Titration Technique: Concentration and Its Uncertainty
  1. 1Colorimetry: the Beer-Lambert law at the bench
  2. 2Colorimetry at work: rates, formulas of complexes and equilibrium
  3. 3

Try the code and the simulations beside the lesson.

Colour and concentration

A coloured solution absorbs some of the light that enters it. The more coloured particles the light meets, the less light comes out. So the light that comes out tells us how much of the coloured substance there is. This is colorimetry.

Transmittance and absorbance

Light of intensity I0I_0I0​ enters the solution and light of intensity III leaves it. The transmittance is the share that gets through:

T=II0\displaystyle T = \dfrac{I}{I_0}T=I0​I​

The absorbance is

A=log⁡10I0I=−log⁡10T\displaystyle A = \log_{10}\dfrac{I_0}{I} = -\log_{10} TA=log10​II0​​=−log10​T

A solution that lets 10 % of the light through has A=1A = 1A=1. One that lets 1 % through has A=2A = 2A=2. Absorbance has no unit.

The Beer-Lambert law

For a dilute solution, the absorbance is proportional to the concentration ccc and to the path length lll of the light in the solution:

A=εcl\displaystyle A = \varepsilon c lA=εcl

The Greek letter ε\varepsilonε (epsilon) is the molar absorption coefficient. With ccc in mol/dm³ and lll in cm, its unit is dm3 mol−1 cm−1\text{dm}^3\ \text{mol}^{-1}\ \text{cm}^{-1}dm3 mol−1 cm−1, which we write L/(mol cm). It depends on the substance and the wavelength. Some books still call it the molar extinction coefficient. The IChO syllabus calls this law the Lambert-Beer law.

Why a logarithm? Each thin layer of the solution absorbs the same share of the light that reaches it. Equal shares in each layer give a fall that is exponential, and the logarithm turns it into a straight line.

The colorimeter

A colorimeter drawn as a row of parts from left to right: a lamp giving white light, a coloured filter, a small square cell of path length l holding the solution, and a detector joined to a meter. Arrows show the light: its intensity is I0 before the cell and a smaller I after it.
  1. A lamp gives white light.
  2. A filter lets through only a narrow band of colours. A spectrophotometer uses a prism or grating instead, and you choose the wavelength.
  3. The light passes through a small cell, the cuvette, usually with l=1.00 cml = 1.00\ \text{cm}l=1.00 cm.
  4. A detector measures III, and the meter shows AAA.

Which filter? A solution looks blue because it absorbs orange and red light. So we use the colour it absorbs most, the complementary colour: an orange or red filter for a blue copper solution, a green filter for purple manganate(VII), a blue green filter for a red iron thiocyanate solution.

The blank. First fill the cuvette with the solvent and any reagents, but not the coloured substance. Set the meter to A=0A = 0A=0. This removes the light lost to the glass and the solvent.

The calibration line

We rarely trust a value of ε\varepsilonε from a book. Instead we make standard solutions of known concentration, measure each one, and plot AAA against ccc.

A calibration graph of absorbance against concentration. Five standard points lie on a straight line through the origin. A dashed line from an absorbance of 0.612 on the vertical axis runs across to the line and down to the concentration of the unknown, about 2.6 × 10⁻⁴ mol/dm³. Above an absorbance of about 1.2 a grey curve bends below the straight line, labelled stray light.

Here is a set of manganate(VII) standards at 525 nm in a 1.00 cm cell.

ccc / 10−410^{-4}10−4 mol/dm³AAA
1.000.236
2.000.473
3.000.705
4.000.944
5.001.178

The best straight line through the origin has a slope of 0.2357 for each 10−410^{-4}10−4 mol/dm³. So ε=2357 L/(mol cm)\varepsilon = 2357\ \text{L/(mol cm)}ε=2357 L/(mol cm). An unknown reads A=0.612A = 0.612A=0.612. We want ccc, so c=A/slopec = A/\text{slope}c=A/slope, which is 2.60×10−4 mol/dm32.60 \times 10^{-4}\ \text{mol/dm}^32.60×10−4 mol/dm3.

Read the unknown only inside the range of the standards. Outside it you are guessing.

When the law fails

The law holds for dilute solutions. At high absorbance the line bends down, for three reasons.

  1. Stray light. A little light reaches the detector without passing through the sample. When the sample lets through very little, this stray light is a large part of what is measured.
  2. Light that is not of one wavelength. A filter passes a band of colours, and ε\varepsilonε is not the same across the band.
  3. Chemical change. At high concentration the particles may pair up or react, and the absorbing species changes.

So a good reading lies between about A=0.1A = 0.1A=0.1 and A=1A = 1A=1. If a reading is higher, dilute the solution and multiply back.

Set 1 % of stray light. The reading stays on the line up to an absorbance near 1. Above 2, the meter cannot read much higher, however strong the solution.

Practise with new numbers

Try it in Python

The program fits the calibration line through the origin and reads the unknown from it.

Read it first. Predict the concentration of the unknown. Then run it.

1. Change the unknown absorbance to 1.05. Is it still inside the range of the standards?

2. Add a sixth standard, 6.50 with an absorbance of 1.53. Does the slope change much?

Common mistakes

  • Using the colour of the solution as the filter. Use the colour it absorbs.
  • Forgetting the blank. Zero the meter with the solvent and reagents first.
  • Reading outside the standards. Dilute instead, and multiply back.
  • Mixing up TTT and AAA. A=−log⁡10TA = -\log_{10} TA=−log10​T, and TTT must be a fraction, not a percentage.

Remember these

  • A=log⁡10(I0/I)=εclA = \log_{10}(I_0/I) = \varepsilon c lA=log10​(I0​/I)=εcl.
  • The filter is the complementary colour.
  • Calibrate with standards, and stay between about A=0.1A = 0.1A=0.1 and A=1A = 1A=1.

References

  • OpenStax Chemistry 2e, Chapter 6, section 6.1 on electromagnetic energy, and Chapter 4 on solution concentration.
  • IUPAC Compendium of Chemical Terminology (the Gold Book), entries on absorbance and on the molar absorption coefficient.

Check yourself

Try each one on your own first. If you are stuck, open a hint. Then check with Show the answer.

  1. A solution lets 25 % of the light through. What is its absorbance?
  1. Why is a green filter used to measure a purple solution of potassium manganate(VII)?
  1. A solution has A=0.30A = 0.30A=0.30 in a 1.00 cm cell. What absorbance does it give in a 2.00 cm cell?
  1. Why should an unknown not be read above the top standard?

Enroll free in Colorimetry and Titration Technique: Concentration and Its Uncertainty

Every lesson is open to read. Enrolling also lets you:

  • Mark lessons done and carry on where you stopped.
  • Take the tests of this lesson and its course.
  • Count the course towards the certificate for its chapter.
No payment. You can leave at any time.

Problems (3)

A student makes up 250.0 cm³ of a dilute solution of potassium manganate(VII), KMnO₄, from a small…Hard · High School (11–12) · by StemPy Lab
Iron(II) ions and 1,10-phenanthroline give an orange red complex that absorbs strongly at 510 nm…Advanced · High School (11–12) · by StemPy Lab
A spectrophotometer lets 1.0 % of stray light reach its detector…Hard · High School (11–12) · by StemPy Lab
Next: lesson 2Colorimetry at work: rates, formulas of complexes and equilibrium
← Colorimetry and Titration Technique: Concentration and Its Uncertainty
← PreviousColorimetry at work: rates, formulas of complexes and equilibrium →

Discussion (0)

Log in to join the discussion.

Loading…

1 Python program
4 questions to check yourself

This lesson is free to read. Enroll free to keep your place and take the tests.

Titration technique, uncertainty and error

Python Labs (1)

A calibration line for a colorimeterPython Lab · by StemPy Lab