Try it on paper first. A hint, the key idea and the worked solution are here when you need them.
A spectrophotometer lets 1.0 % of stray light reach its detector. A strongly coloured solution has a true absorbance, by the Beer-Lambert law, of 1.50. A second, weaker solution has a true absorbance of 0.50. Use the stray light formula in the reminder. Find the absorbance the instrument reads for each solution. Then find the percentage by which the concentration of the strong solution would be underestimated, if the reading were used with the true calibration slope. Then advise the analyst.
Check an answer or open a hint to unlock the worked solution.
Have you tried it?
Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.
Step 1. The strong solution. . The meter reads .
Step 2. The weak solution. . It reads .
Step 3. The error. For the strong solution, too low. For the weak one it is only .
Answer: , and .
Advice. Dilute the strong solution, for example by a factor of 3, so that its absorbance falls below 1. Then multiply the concentration back by 3.
Watch out. Stray light matters most when the sample lets little light through. That is why the line bends at high absorbance and not at low absorbance.
Log in to join the discussion.
Loading…
Absorbance , and the Beer-Lambert law is . A calibration line of against gives the concentration of an unknown. With a fraction of stray light, the meter reads .