An NDIR analyser samples the air in a road tunnel during a traffic jam. Its sample cell is long, and for carbon monoxide per ppm per metre. The detector shows that of the infrared light gets through the sample cell, compared with the reference cell. (a) What is the concentration of carbon monoxide in ppm? (b) What is it in milligrams per cubic metre at ? The molar mass of carbon monoxide is . (c) What share of the light would get through at ? (d) Why does the analyser need a reference cell?
Check an answer or open a hint to unlock the worked solution.
Have you tried it?
Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.
Step 1. The concentration. .
Step 2. Mass units. .
Step 3. At 10 ppm. .
Step 4. The reference cell. The source can grow brighter or dimmer, and the detector can drift. Comparing the sample beam with a beam through gas that does not absorb cancels these changes, so only the absorption by carbon monoxide is left.
Answer: ; ; ; to cancel changes in the source and the detector.
Check. At less than of the light is absorbed. NDIR analysers for clean air therefore use long paths and very steady detectors.
Log in to join the discussion.
Loading…
An NDIR analyser follows the Beer-Lambert law, , so . At and 1 atm, a concentration in ppm becomes milligrams per cubic metre on multiplying by , with the molar mass in grams per mole.