Sampling the air: stations, samplers and analysers
We can measure pollution where it leaves a chimney or in the open air that people breathe. A network of monitoring stations needs the right places, types, numbers and timing. Manual methods catch the pollutant on a filter or in a liquid; instruments measure it with light, as NDIR and chemiluminescence do.
about 35 min1 simulation1 practice question
Lesson 3 of 3Show all lessonsHide
Try the code and the simulations beside the lesson.
Inside a monitoring station
Climb onto the roof of a monitoring station and you see a few pipes pointing at the sky and a box like a large kitchen cabinet. One pipe leads to a pump that pulls air through a filter. Others lead to analysers that glow and click all day. A screen shows the levels of six gases, updated every few minutes. How does each part work, and why is the station here and not one street away?
Two places to measure
Monitoring at the source measures what one chimney or one exhaust pipe gives out. A probe is put into the stack, and the gas and dust are sampled as they flow. The result, in kilograms an hour or milligrams per cubic metre of flue gas, is checked against the emission standard of that industry.
Ambient air quality monitoring measures the outdoor air that people breathe. It shows the effect of all the sources together, after the wind has mixed and carried them. The result is checked against the ambient air quality standards.
A factory can meet its emission limit and the town can still have poor air, because many sources add up. So we need both kinds.
Where to put a station
A station must give data that stand for a whole area. The main criteria for choosing a site are these.
Representative. The site shows the air of the area, not of one nearby chimney or kitchen.
Free flow of air. No tall building, wall or tree close to the inlet. The inlet is usually 3 to 10 m above the ground.
Purpose. Downwind of an industrial area to measure its effect, or in a crowded area to measure what people breathe.
Practical. A secure place with power, easy access for the staff, and use of the site for many years.
Types of stations
Stations are named by the area they stand for:
residential, where people live;
industrial, near factories;
commercial or traffic, beside busy roads and markets;
sensitive, near hospitals, schools or protected forests;
background, far from sources, to show the clean level of a region.
Stations also differ in how they work. A manual station collects samples that are taken to a laboratory. A continuous station has analysers that give real-time data all day.
How many stations, and how often
Number of stations. A large city with many sources needs more stations than a small town. The number depends on the area and the population, and on how many sources there are and how spread out. It also depends on how much the pollution changes from place to place, and on the money available. Each type of area should have at least one.
Frequency of data. Continuous analysers give a reading every few minutes, averaged into hourly values. At a manual station under the National Air Quality Monitoring Programme, each pollutant is sampled for 24 hours, 2 times a week, which gives 104 samples a year. Gases are sampled in 4-hour parts of the day and particles in 8-hour parts.
Sampling the ambient air, in brief
A good sample of air follows a few basic rules.
It stands for the place and the time it is meant to describe.
It has a known volume of air, which is the flow rate times the time. So the flow is checked and the pump is calibrated.
Its averaging time matches the standard it is compared with: 1 hour, 8 hours, 24 hours or a year.
The pollutant is caught completely, and it does not change before it is analysed.
The weather is recorded: wind, temperature, pressure.
Manual techniques
Passive samplers have no pump. A small tube or badge holds a material that absorbs one gas. The gas reaches it by diffusion, at a known rate. After one to four weeks the sampler goes to a laboratory. Passive samplers are cheap and silent, so many can be placed across a city. They give only an average over the weeks, not the peaks of one day.
High volume samplers measure particles. A strong pump pulls air at about 1.1 to 1.7m3 a minute through a glass fibre filter. The filter is weighed before and after sampling, usually for 8 or 24 hours. A respirable dust sampler adds a cyclone that throws out the larger particles, so the filter catches only the PM10.
concentration=flow rate×timemass gained by the filter
Bubbler systems measure gases. A small pump bubbles air slowly through a liquid that absorbs one gas: sulphur dioxide in one bottle, nitrogen dioxide in another. In the laboratory a reagent turns the absorbed gas into a coloured compound. The depth of the colour, read with a colorimeter, gives the mass of the gas. Dividing by the volume of air gives the concentration.
Instrumental techniques
Analysers measure the gas as it flows, using light. They are called photometric techniques.
NDIR (non-dispersive infrared) measures carbon monoxide and carbon dioxide. Each of these gases absorbs infrared light of its own wavelengths: carbon monoxide near 4.6μm. The analyser shines infrared light through a cell of sample air and through a reference cell. A filter picks out the wavelength of the gas, so no prism or grating is needed. That is what "non-dispersive" means. The less light that gets through the sample cell, the more of the gas it holds.
A diagram of a non-dispersive infrared analyser for carbon monoxide, drawn from left to right. An infrared source sends two beams through a rotating chopper. The upper beam passes through a sample cell through which the air being tested flows in and out. The lower beam passes through a reference cell filled with a gas that does not absorb, such as nitrogen. Both beams then pass through an optical filter that lets through only light near 4.6 micrometres, the wavelength that carbon monoxide absorbs, and reach a detector. The detector compares the two beams: the weaker the sample beam, the more carbon monoxide.
The light that gets through falls off exponentially with the amount of gas, by the Beer-Lambert law:
T=e−kcL
Here T is the share of light that gets through, k the absorption coefficient, c the concentration and L the length of the cell.
Chemiluminescence measures nitrogen oxides. In the analyser, nitric oxide in the sample meets ozone made inside the instrument:
NO+O3→NO2∗+O2
NO2∗→NO2+light
The star marks a molecule with extra energy. It gives this energy off as a faint glow of red and near-infrared light. A sensitive detector, a photomultiplier, measures the glow, which is proportional to the nitric oxide. To measure nitrogen dioxide, part of the sample first passes over a hot converter that turns NO2 into NO. That channel gives all the NOx, and NO2=NOx−NO.
Worked example: a filter and an analyser
A high volume sampler runs at 1.3m3/min for 8 hours, and its filter gains 87.4mg.
Volume of air.1.3m3/min×8×60min=624m3.
Mass.87.4mg=87400μg.
Concentration.87400μg/624m3=140μg/m3.
An NDIR analyser has a cell 0.5m long, and k=0.0009 per ppm per metre for carbon monoxide. Only 99.6% of the light gets through.
Rearrange. We want c, so c=−lnT/(kL).
Put in the numbers.c=−ln0.996/(0.0009×0.5)=8.9ppm.
In mass units. At 25∘C and 1 atm a mole of gas fills 24.45L. So 8.9ppm of carbon monoxide is 8.9×28.01/24.45=10.2mg/m3.
With the sliders as they are, what concentration does a reading of 0.98 mean? Now halve the length of the cell. Is the same small amount of gas easier or harder to tell apart from clean air?
Practise with new numbers
Try it in Python
The program reads the two channels of a chemiluminescence analyser, the NO channel and the NOx channel, and works out the nitrogen dioxide for each hour.
Common mistakes
Mixing up source and ambient monitoring. A probe in a chimney measures one source. A station in a park measures the air people breathe.
Forgetting to turn hours into minutes. The flow of a high volume sampler is given per minute.
Thinking that a passive sampler shows the worst hour. It gives an average over days or weeks.
Saying that chemiluminescence measures nitrogen dioxide directly. It measures nitric oxide. Nitrogen dioxide is found by converting it to nitric oxide first.
Remember these
Source monitoring: one chimney, against emission standards. Ambient monitoring: the outdoor air, against air quality standards.
Stations: representative site, free air flow, inlet 3 to 10 m up; residential, industrial, traffic, sensitive and background.
Manual sampling: 24 hours, 2 times a week, 104 samples a year. Continuous analysers: real-time data.
Manual: passive samplers, high volume samplers (particles on a filter), bubblers (gases in a liquid).
Instrumental: NDIR for CO and CO2, with T=e−kcL; chemiluminescence for NO and NOx.
Check yourself
Try each one on your own first. If you are stuck, open a hint. Then check with Show the answer.
Why is a station not put right next to a tall wall or a roadside food stall?
Which technique would you use to map nitrogen dioxide across fifty streets cheaply, over one month?
A high volume sampler pulls 1.5m3 of air a minute for 2 hours. How much air passes through it?
What makes the glow in a chemiluminescence analyser?
References
Central Pollution Control Board, Guidelines for the Measurement of Ambient Air Pollutants, Volume I (manual sampling and analysis).
Central Pollution Control Board, National Air Quality Monitoring Programme, the pages on objectives and methods.
D. A. Skoog, F. J. Holler and S. R. Crouch, Principles of Instrumental Analysis, the chapters on infrared and molecular luminescence.
Enroll free in Air Pollution: Sources, Control and Sampling
Every lesson is open to read. Enrolling also lets you:
Mark lessons done and carry on where you stopped.
Take the tests of this lesson and its course.
Count the course towards the certificate for its chapter.