A brief note on Nuclear Medicine, Dose and Radiotherapy . Read it before a test. Try each question at the end before you open its answer.
Tracers and the gamma camera A good tracer: a gamma emitter of about 100 100 100 to 200 200 200 keV, a half-life of hours, joined to the right chemical. Technetium-99m: 140 140 140 keV gamma rays, half-life 6.01 6.01 6.01 h, made in a generator. Activity: A = λ N A = \lambda N A = λ N , with λ = ln 2 / T \lambda = \ln 2/T λ = ln 2/ T . Effective half-life: 1 T e = 1 T p + 1 T b \dfrac{1}{T_e} = \dfrac{1}{T_p} + \dfrac{1}{T_b} T e 1 = T p 1 + T b 1 . Gamma camera: lead collimator, sodium iodide crystal, photomultiplier tubes, pulse height to reject scatter. PET Tracer: FDG with fluorine-18, half-life 110 110 110 min, taken up by cells that use glucose fast. Annihilation: e + + e − → 2 γ e^+ + e^- \to 2\gamma e + + e − → 2 γ , each photon 0.511 0.511 0.511 MeV, in opposite directions. Coincidence: two detectors that fire together give a line of response. Time of flight: x = c Δ t 2 x = \dfrac{c\,\Delta t}{2} x = 2 c Δ t from the midpoint. Dose and therapy Absorbed dose: D = E / m D = E/m D = E / m , in gray ( 1 Gy = 1 J/kg 1\ \text{Gy} = 1\ \text{J/kg} 1 Gy = 1 J/kg ). Equivalent dose: H = w R D H = w_RD H = w R D , in sievert. w R = 1 w_R = 1 w R = 1 for photons, 2 2 2 for protons, 20 20 20 for alpha. Effective dose: E = ∑ w T H T E = \sum w_TH_T E = ∑ w T H T . The tissue factors add up to 1 1 1 . Risk: about 5 % 5\ \% 5 % per sievert. Background is about 2.4 2.4 2.4 mSv a year. Protection: time, distance ( 1 / r 2 1/r^2 1/ r 2 ) and shielding. Radiotherapy: cross-fire, daily fractions, brachytherapy, iodine-131, protons with a Bragg peak. A worked example A tracer has T p = 6.0 T_p = 6.0 T p = 6.0 h and T b = 30 T_b = 30 T b = 30 h. Find the effective half-life.
The rates. 1 T e = 1 6.0 h + 1 30 h = 6 30 h \dfrac{1}{T_e} = \dfrac{1}{6.0\ \text{h}} + \dfrac{1}{30\ \text{h}} = \dfrac{6}{30\ \text{h}} T e 1 = 6.0 h 1 + 30 h 1 = 30 h 6 . The answer. T e = 5.0 T_e = 5.0 T e = 5.0 h, shorter than both. Practise with new numbers
Check yourself Why does PET need two detectors to fire at the same time? Stuck? Show a hint Show the answerA tissue absorbs 2.0 2.0 2.0 mGy of gamma rays. What is its equivalent dose? Stuck? Show a hint Show the answerWhy is a gamma emitter preferred to a beta emitter for imaging? Stuck? Show a hint Show the answerSources OpenStax College Physics 2e, Chapter 32 Medical Applications of Nuclear Physics. ICRP Publication 103 (2007), for the weighting factors. Cambridge International AS and A Level Physics (9702) syllabus, Medical physics, PET scanning.