A brief note on X-Ray Imaging, CT and MRI . Read it before a test. Try each question at the end before you open its answer.
The X-ray tube Parts: hot cathode, vacuum, accelerating voltage of 20 20 20 to 150 150 150 kV, spinning tungsten anode, lead housing. Spectrum: continuous braking radiation, with sharp lines of the target on top. Shortest wavelength: λ m i n = h c e V \lambda_{min} = \dfrac{hc}{eV} λ min = e V h c , with h c / e = 1239.8 hc/e = 1239.8 h c / e = 1239.8 nm V. Energy: only about 1 % 1\ \% 1 % becomes X-rays. The rest heats the anode. Current: sets the number of photons. Voltage: sets their largest energy. Filter: aluminium removes soft photons and hardens the beam. Attenuation and contrast Law: I = I 0 e − μ x I = I_0e^{-\mu x} I = I 0 e − μx . Layers add in the exponent. Half-value thickness: x 1 / 2 = ln 2 / μ x_{1/2} = \ln 2/\mu x 1/2 = ln 2/ μ . Mass coefficient μ m = μ / ρ \mu_m = \mu/\rho μ m = μ / ρ . Photoelectric effect: about Z 3 / E 3 Z^3/E^3 Z 3 / E 3 , the main source of contrast. Compton scattering: main process in tissue above about 30 30 30 keV. It fogs the picture. Contrast media: barium ( Z = 56 Z = 56 Z = 56 ) for the gut, iodine ( Z = 53 Z = 53 Z = 53 ) for vessels. Sharpness: small focal spot, grid against scatter, short exposure. CT and MRI CT: a fan beam turns around the body, and a computer finds each voxel. CT number: 1000 ( μ − μ w ) / μ w 1000(\mu - \mu_w)/\mu_w 1000 ( μ − μ w ) / μ w . Water 0 0 0 HU, air about − 1000 -1000 − 1000 HU. MRI: f = 42.58 MHz/T × B f = 42.58\ \text{MHz/T} \times B f = 42.58 MHz/T × B for hydrogen. A pulse at f f f tips the protons. Contrast in MRI: different relaxation times. Position: a gradient in B B B . Safety: CT gives a much larger dose than a plain X-ray. MRI gives none, but metal must be kept out. A worked example A tube runs at 90 90 90 kV. Find the shortest wavelength and the largest photon energy.
The energy. One electron brings 90 90 90 keV, so no photon has more than 90 90 90 keV. The wavelength. λ m i n = 1239.8 nm V 90 000 V = 0.0138 \lambda_{min} = \dfrac{1239.8\ \text{nm V}}{90\,000\ \text{V}} = 0.0138 λ min = 90 000 V 1239.8 nm V = 0.0138 nm, which is 13.8 13.8 13.8 pm. Practise with new numbers
Check yourself A lead sheet has a half-value thickness of 0.20 0.20 0.20 mm. What share of a beam passes 0.80 0.80 0.80 mm? Stuck? Show a hint Show the answerWhy does a scanner for MRI need a gradient field as well as the main field? Stuck? Show a hint Show the answerWhy is the anode of a tube made to spin? Stuck? Show a hint Show the answerSources OpenStax College Physics 2e, Chapter 30 Atomic Physics (X-rays) and Chapter 32 Medical Applications of Nuclear Physics. Cambridge International AS and A Level Physics (9702) syllabus, Medical physics, production and use of X-rays.