In a PET scanner, one photon of a pair reaches its detector 0.64 ns before the other photon reaches the opposite detector. How far from the midpoint of the line of response was the annihilation?
Physics
Hard
High School (11–12)
Medical Physics
A-Level
Created by
StemPy Lab
In a PET scanner, one photon of a pair reaches its detector
0.64
0.64
0.64
ns before the other photon reaches the opposite detector. How far from the midpoint of the line of response was the annihilation?
0.959
cm
0.959\ \text{cm}
0.959
cm
4.80
cm
4.80\ \text{cm}
4.80
cm
9.59
cm
9.59\ \text{cm}
9.59
cm
19.2
cm
19.2\ \text{cm}
19.2
cm
Check Answer
Log in to save
Tag someone
⚑ Report
← All Quizzes
← A patient is given 550 MBq of iodine-131, whose half-life is 8.02 days. Take its molar mass as 131 g/mol and NA = 6.022 × 10²³ mol⁻¹. What mass of iodine-131 is that?
A student writes: when a positron and an electron at rest annihilate, each of the two photons carries 1.022 MeV. What is wrong? →
More on
Medical Physics
Discussion (0)
Log in
to join the discussion.
Loading…
All questions on this topic →
Problems on this topic
A tumour of mass 0.12 kg is to receive 60 Gy in 30 equal fractions…
Advanced
After a large dose of iodine-131, a patient gives a dose rate of 25 microsievert per hour at 1.0 m…
Hard
At the start of a scan, a patient holds 300 MBq of fluorine-18, whose half-life is 109.8 min…
Advanced
A cyclotron makes fluorine-18, with a half-life of 109.8 min, for a hospital 2.0 h away…
Hard
All problems on this topic →