HelpContactFor teachers and institutes
>_StemPy Lab
Mock TestsClasses

>_StemPy Lab

Learn. Solve. Code the solutions. Test yourself.

Live examsFree mock exams, timed and rankedEveryone starts together, and your score and rank come the moment it ends.See the exams →
Live examsInternational live exams are comingTell us which exam you want first. Timed practice papers are open now.See what is coming →

Learn

  • Physics
  • Chemistry
  • Mathematics
  • Biology
  • Python Programming
  • Statistics & Data Science
  • Computer Science
  • Space Science and Astronomy
  • Classes
  • Groups

Practice

  • Problems
  • Quizzes
  • Mock Tests
  • Prepare for an Exam
  • Question Sets
  • Exams
  • Certificates
  • Python Labs
  • AI Practice Generator
  • Student-Created Tests

Community

  • Discussions
  • Resources
  • Notes
  • Glossary
  • Blog

Account

  • Log in
© 2026 senLearn. All rights reserved.HelpFor SchoolsAboutContactPrivacy PolicyTerms of UseRefund Policy
  1. Problems
  2. /The holes of a lead collimator are 1.5 mm wide and 25 mm long…
AdvancedPhysicsHigh School (11–12)Medical PhysicsNumericalConceptualA-Level

The holes of a lead collimator are 1.5 mm wide and 25 mm long…

StemPy LabContributed by StemPy Lab

Try it on paper first. A hint, the key idea and the worked solution are here when you need them.

The holes of a lead collimator are 1.51.51.5 mm wide and 252525 mm long. A gamma ray can pass a hole at a small angle, as long as it enters at one edge and leaves at the other. So a point source at a distance zzz in front of the collimator is seen as a blurred spot. Its width is about d(L+z)/Ld(L + z)/Ld(L+z)/L, where ddd is the width and LLL the length of a hole. Find the width of the spot for a source 100100100 mm in front of the collimator, and for one 202020 mm in front. A second collimator has holes 404040 mm long. Find the width for the source at 100100100 mm. What does the longer collimator cost?

Check your answer

Work it out on paper, then type your number. The answer itself stays hidden.

mm
mm
mm
Need a reminder? The key ideaShowHide

The activity is A=λNA = \lambda NA=λN with λ=ln⁡2/T\lambda = \ln 2/Tλ=ln2/T, and it falls as A=A0×2−t/TA = A_0 \times 2^{-t/T}A=A0​×2−t/T. When the body also removes the tracer,

1Te=1Tp+1Tb\displaystyle \frac{1}{T_e} = \frac{1}{T_p} + \frac{1}{T_b}Te​1​=Tp​1​+Tb​1​

The worked solution

Check an answer or open a hint to unlock the worked solution.

Have you tried it?

Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.

Step 1. At 100 mm. 1.5 mm×(25 mm+100 mm)25 mm=7.5\dfrac{1.5\ \text{mm} \times (25\ \text{mm} + 100\ \text{mm})}{25\ \text{mm}} = 7.525 mm1.5 mm×(25 mm+100 mm)​=7.5 mm.

Step 2. At 20 mm. 1.5 mm×(25 mm+20 mm)25 mm=2.7\dfrac{1.5\ \text{mm} \times (25\ \text{mm} + 20\ \text{mm})}{25\ \text{mm}} = 2.725 mm1.5 mm×(25 mm+20 mm)​=2.7 mm.

Step 3. Longer holes. 1.5 mm×(40 mm+100 mm)40 mm=5.25\dfrac{1.5\ \text{mm} \times (40\ \text{mm} + 100\ \text{mm})}{40\ \text{mm}} = 5.2540 mm1.5 mm×(40 mm+100 mm)​=5.25 mm.

Answer: 7.57.57.5 mm, 2.72.72.7 mm and 5.255.255.25 mm.

The cost. Longer holes accept a narrower cone of rays, so fewer gamma rays get through. The picture is sharper, but it takes longer to collect, or needs more activity. This is why the camera is brought as close to the patient as it can go.

Log in to save this for later
← All Problems
← A patient with an overactive thyroid is given iodine-131…A hospital must choose a tracer for a kidney scan that takes 4.0 h from injection to the last… →
More onMedical Physics

Discussion (0)

Log in to join the discussion.

Loading…

Learn it in

  1. Radioactive tracers and the gamma cameraNuclear Medicine, Dose and Radiotherapy

More on this topic

  1. A tumour of mass 0.12 kg is to receive 60 Gy in 30 equal fractions…Advanced
  2. After a large dose of iodine-131, a patient gives a dose rate of 25 microsievert per hour at 1.0 m…Hard
  3. At the start of a scan, a patient holds 300 MBq of fluorine-18, whose half-life is 109.8 min…Advanced
A cyclotron makes fluorine-18, with a half-life of 109.8 min, for a hospital 2.0 h away…Hard
  • A PET ring has a diameter of 80 cm. For one event, the photon that reaches detector A arrives 300…Hard
  • All problems on this topic →