Galvanometer, ammeter and voltmeter
A moving-coil galvanometer has a coil in a radial field, so the torque NIAB balances the spring at kϕ, and ϕ=kNABI
Lesson 2 of 3Show all lessonsHide
Try the code and the simulations beside the lesson.
How does a meter know the current?
Look at the pointer of an old meter on a workshop bench. The needle swings further when more current flows. How does the current move the needle?
It uses the torque of the last lesson. A coil with a pointer sits between two magnets. The current in the coil gets a torque. A spring pulls back. The needle stops where the two torques are equal. The instrument is a moving-coil galvanometer.
A radial field
In a uniform field, the torque NIABsinθ changes as the coil turns. The scale would be uneven. Worse, the torque would die away as the coil swings to the field.
The fix is to shape the field. Curved pole pieces surround a fixed cylinder of soft iron. In the narrow gap, the field points along the radius, like the spokes of a wheel. This is a radial field.
Two panels. The left panel is a top view of the inside of a moving-coil galvanometer. A curved north pole piece on the left and a curved south pole piece on the right face each other across a gap, with a grey soft iron cylinder at the centre. Short blue arrows run from the left pole across the gap into the cylinder and out again to the right pole, all pointing along the radius of the cylinder: a radial field. Two small circles in the gap, one with a dot and one with a cross, show the two sides of the coil, which lie in the radial field wherever the coil turns. The right panel is a graph of torque against the angle of deflection. A straight rising line through the origin is the torque of the spring. Two horizontal lines, labelled for a smaller and a larger current, are the torque of the magnetic force in a radial field, which stays the same at every angle. Each horizontal line meets the spring line at a point marked on the angle axis, and the larger current gives the larger angle, in proportion. A grey dashed curve that falls as the angle grows shows the magnetic torque in a uniform field for comparison.
The coil turns in the gap. Wherever it turns, its plane contains the field, so the normal is at 90∘ to the field. The torque is always
τmagnetic=NIAB
A spring (or a pair of hair-springs) pulls back with a torque kϕ when the coil has turned through an angle ϕ. Here k is the torsion constant of the spring, in N m/rad. The needle rests when the two are equal:
NIAB=kϕsoϕ=kNABI
Look at the result. The deflection is in direct proportion to the current. That is why the scale of a galvanometer has even divisions. In the graph, the magnetic torque is a flat line, the spring torque is a rising line, and a bigger current moves the meeting point out in proportion.
Sensitivity
A galvanometer is judged by how far the needle moves for a small current.
Current sensitivity:Iϕ=kNAB. It is the deflection for each unit of current.
Voltage sensitivity:Vϕ=kGNAB, where G is the resistance of the coil. It is the deflection for each unit of voltage.
To raise the current sensitivity, use more turns, a larger area, a stronger magnet and a weaker spring. But take care with the voltage sensitivity. Doubling N doubles NAB/k. It also doubles the length of wire, so it doubles G. The voltage sensitivity then stays the same.
A worked example. A coil has 100 turns of area 5.0cm2 in a radial field of 0.16 T. The spring has k=1.6×10−6N m/rad. Find the deflection for 60μA.
The factor.NAB=100×5.0×10−4m2×0.16T=8.0×10−3N m/A.
The sensitivity.Iϕ=1.6×10−6N m/rad8.0×10−3=5000rad/A.
The deflection.ϕ=5000rad/A×60×10−6A=0.30rad, which is 17∘.
Too delicate for the real world
A galvanometer is made to be sensitive. Its full-scale current Ig is only a fraction of a milliampere or a few milliampere. Its coil has a resistance G of tens of ohms. Send a whole ampere through it and the coil burns. A voltage of a few volt across it does the same.
So we cannot use it as it is to measure amperes or volts. We add a resistor to it. The resistor makes the meter fit the job.
An ammeter: add a shunt
An ammeter measures a current, so it is joined in series with the part of the circuit that carries that current. It must not change the current. So it needs a very small resistance.
We join a small resistor, the shuntS, in parallel with the coil. Most of the current goes through the shunt. The coil gets only Ig when the circuit current is I, the range we want.
The coil and the shunt have the same voltage across them:
IgG=(I−Ig)S
We want S, so
S=I−IgIgG
The ammeter is the coil and the shunt together, in parallel. At full scale the voltage across it is IgG and the current into it is I. So its resistance is RA=IIgG, which is smaller than both G and S.
A worked example. The coil has G=80Ω and Ig=1.5 mA. We want an ammeter for 0 to 3.0 A.
The shunt.S=3.0A−1.5×10−3A1.5×10−3A×80Ω=0.0400Ω=40.0mΩ.
The meter.RA=IIgG=3.0A1.5×10−3A×80Ω=40.0mΩ.
The scale is then marked again. The deflection that meant 1.5 mA now means 3.0 A.
Raise the range of the ammeter. Does the shunt get larger or smaller? Now raise the resistance of the coil. What happens to the shunt?
A voltmeter: add a resistance in series
A voltmeter measures a voltage, so it is joined in parallel across the part that is measured. It must not take much current from the circuit. So it needs a very large resistance.
We join a large resistor R in series with the coil. At full scale, the voltage V is across the resistor and the coil together:
V=Ig(G+R)
We want R, so
R=IgV−G
The voltmeter is the coil and the resistor in series, so its resistance is RV=G+R=IgV.
A worked example. The same coil, G=80Ω and Ig=1.5 mA, is to read 0 to 9.0 V.
The resistor.R=1.5×10−3A9.0V−80Ω=6000Ω−80Ω=5920Ω.
The meter.RV=IgV=6000Ω. Its rating is Ig1=667Ω for each volt of range. The higher the rating, the better the voltmeter.
Two circuit diagrams. In the left diagram, an ammeter: a current I enters from the left, divides at a junction into two branches and joins again. The upper branch holds a circle marked G, the galvanometer, with the small current I g in it. The lower branch holds a zigzag resistor marked S, the shunt, with the current I minus I g in it. A note says the ammeter is joined in series with the circuit, has a very small resistance and is read in amperes. In the right diagram, a voltmeter: a circle marked G in series with a zigzag resistor marked R, between two terminals. A note says the voltmeter is joined across the part of the circuit that is measured, has a very large resistance and is read in volts.
Meters change the circuit
A meter is part of the circuit it measures, so it changes the circuit a little.
An ammeter adds its resistance in series. The current falls a little. The reading is a little low.
A voltmeter in parallel takes a current. The resistance of that part of the circuit falls, so the voltage across it falls. The reading is a little low.
An ideal ammeter has zero resistance. An ideal voltmeter has infinite resistance. No real meter is ideal. A real voltmeter should have a resistance far larger than the resistance it is joined across. The graph shows how the error falls as the resistance of the voltmeter grows.
Practise with new numbers
Try it in Python
The program uses one coil to make four ammeters and three voltmeters. It prints the shunt, the series resistor and the resistance of each meter.
Common mistakes
Joining the shunt in series. A shunt is always in parallel with the coil.
Using I in place of I−Ig. The shunt carries what the coil does not take.
Forgetting to subtract G. For a voltmeter, R=V/Ig−G.
Joining an ammeter in parallel across a part of a circuit. Its tiny resistance takes a huge current.
Giving the voltmeter a small resistance. A voltmeter wants a large resistance and an ammeter a small one.
Mixing mA and A. Change to A before you divide.
Thinking that more turns always give a better voltage sensitivity. The resistance of the coil grows with the turns too.
Remember these
Radial field: the torque is NIAB at every angle. Deflection: ϕ=kNABI.
Current sensitivity kNAB. Voltage sensitivity kGNAB.
Ammeter: shunt in parallel, S=I−IgIgG. Low resistance, joined in series.
Voltmeter: resistance in series, R=IgV−G. High resistance, joined in parallel.
An ideal ammeter has zero resistance. An ideal voltmeter has infinite resistance.
Check yourself
Try each one on your own first. If you are stuck, open a hint. Then check with Show the answer.
A galvanometer has a full-scale current of 0.25 mA and a coil resistance of 100Ω. What voltage is across the coil at full scale?
The same galvanometer is to read 0 to 5.0 V. What resistance must be joined in series?
Why must an ammeter have a small resistance and a voltmeter a large one?
A student joins an ammeter across a resistor by mistake, in parallel. What happens?