A chemist runs one sample on two spectrometers. On the 300 MHz machine a methoxy signal lies 1092 Hz from TMS. Two other signals of the sample, a triplet and a quartet with J = 7.0 Hz, have shifts that differ by 0.10 ppm. Find the chemical shift of the methoxy signal. Then find how many hertz it lies from TMS on the 600 MHz machine. Last, find how many hertz separate the centres of the two multiplets on the 600 MHz machine, and say why the spectrum looks cleaner there.
(protons , carbon-13 ), . The ratio of the upper to the lower population is . : the offset in hertz is times the spectrometer frequency in MHz.
Check an answer or open a hint to unlock the worked solution.
Have you tried it?
Work it out on paper first. It is fine to get stuck or make a mistake: that is how we learn. When you are ready, go through the solution one step at a time, and check each step against your own.
Step 1. The shift. .
Step 2. At 600 MHz. .
Step 3. The multiplets. apart, against 30 Hz at 300 MHz.
Answer: , and .
Why cleaner. Each multiplet is still 7.0 Hz between lines, so a triplet spans 14 Hz and a quartet 21 Hz. At 300 MHz the gap between their nearest outer lines is . At 600 MHz it is , so the multiplets stand well apart and the n + 1 rule works better.